High-frequency electric knife operation instructions
A high-frequency electrocution is a medical device that produces thermal effects through high-frequency currents to perform tissue cutting, coagulation or ablation. The following is the instruction manual for the high-frequency electric knife:
一、Preparation before use
1. Check the equipment: Make sure that the host, foot switch, power cord and electrode accessories are not damaged and the connection is normal.
2. Install electrodes: Select a monopole/bipolar mode according to surgical needs. The monopole mode needs to be attached to the negative electrode plate (placed in the patient's muscle-rich parts, avoiding the bone protrusions and metal implants).
3. Disinfection preparation: Connect the sterilized electric blade to the host machine, check the cleanliness of the knife head to ensure no tissue residue.
二、Power-on setting
1. Turn on the power supply: connect to the 220V power supply, turn on the host switch, and complete the self-test program.
2. Mode selection: Select cutting (Cut) or coagulation (Coag) mode according to the tissue type, and mixing mode (Blend) is recommended for gastrointestinal surgery.
3. Power regulation: Refer to the standard setting (general cutting power is 40-60W, coagulation power is 60-80W), and obese patients can appropriately increase by 10-15%.
三、Intraoperative operation
1. Activate the device: Activate through the foot switch or handle button to avoid no-load activation before contacting the tissue.
2. Cutting skills: Keep the cutting head in contact with the tissue at an angle of 15-30°, move at a constant speed, and avoid long-term stay (single point operation
3. Hemostatic treatment: Use point-to-touch operation when coagulation, and the best effect is 1-2mm away from the bleeding point.
四、 Postoperative treatment
1. Turn off the equipment: Turn off the mains power first and then disconnect the power cord.
2. Cleaning and disinfection: Remove the cutter head and use a soft cloth to remove the eschar and wipe the body with a special disinfectant.
Notes:
1. It is prohibited to use in environments with flammable anesthetic gases
2. Patients installing a pacemaker should use it with caution
3. The contact area of the negative plate must be >100cm²
4. The equipment temperature should be checked if it is used continuously for more than 1 hour.
The high-frequency ablation electrosurgery produced by Simon Medical uses high-tech technology innovative products that combine high-frequency electrosurgery and low-temperature ablation, which has safe treatment and low recurrence rate. Welcome to call: 13820671842 for business consultation
Note: There may be differences in high-frequency electric knives of different brands/models. Please be sure to read the equipment special instruction manual before operation!